RIGOL  Chapter 2 Front Panel Operations 
2-30   DSG800 User Guide 
2.  External Source 
When "Ext" is selected, the RF signal generator receives the in-phase baseband 
signal and quadrature phase modulating signal of I/Q modulation input from the 
[I IN] and [Q IN] connectors (as shown in the figures below) at the rear panel 
respectively.   
 
   
 
Tip: When an external baseband signal is input and the baseband output is 
turned "On", the instrument can output the I (In-Phase) components and Q 
(Quadrature Phase) components of the I/Q modulation baseband signal 
generated by the built-in baseband generator (wavetable) from the [I OUT] 
and [Q OUT] connectors at the rear panel respectively. 
 
 
Baseband Signal 
 
1.  Baseband Output 
 
Press BaseBand  Switch to select "On" or "Off". 
  On: turn on the baseband output function. The RF signal generator outputs 
the I (In-Phase) and Q (Quadrature Phase) components of the I/Q 
modulation baseband signal from the [I OUT] and [Q OUT] connectors at 
the rear panel respectively. 
  Off: turn off the baseband output function. 
 
2.  Baseband Level 
 
Press BaseBand  Level to set the amplitude of the baseband output signal. 
  Use the numeric keyboard to input the amplitude value and select the 
desired unit from the pop-up unit menu. 
  The amplitude units available are V, mV, uV, nV and dBm. 
  You can press Backspace to delete the number at the left of the cursor. 
  You can press the left/right arrow key to enter the parameter editing state 
and move the cursor to the specified digit; then, rotate the knob to modify 
the parameter value.   
  You can also rotate knob to modify the amplitude at the current step.   
  You can press Step to set the step. 
 
3.  ARB 
 