PT0E‑0730
12
3−6
.Operationalprocedure
Approachthesamplinghosetothesamplingsiteformeasurementandleavesamplinggas
sucked.
・Takecarethatanytwistonthesamplinghosecannotbemade.
・Stopsamplinghoseatthepointofmeasurement.
・Fromthehoselengthconsiderthedelaytimeandreadoutthereadingafterthereading
getsstable.
Danger
・Oxygendepletionairetcmaybeexhaustedfromgasoutlet.Donotbreatheinthere
absolutely.
・Highdensitygasover(LEL)maybeexhausted.Donotapproachfirethereabsolutely.
Caution
・Whensuckaircontaininggas,doitafterremovingwatervapours.
・Whensuckhightempgas,doitafterbeingusedtobeambienttemp.
・Donothavewaterandoilsuckedwrongly,theinternalpumpandsensormaybetroubled.
・Whenmeasure,checktheroaringsoundofpumpandflowmonitor.
Itisimpossibletomeasureatpumpworkstop.
・Donotblockthepumpoutlet,thereadingshallshowalittlehigher.
・Whenmeasureoxygenininertgasforalongtime,besuretomeasurelowerthan15%CO2
byallmeans.Whenexeedhigherdensitythan15%,shortenthetimeforthismeasurement
aslifeofoxygensensormaybeshortend.
Warning
・Intheplacewhereenvironmenttempisremarkablychanged,theremaybethecasethat
accurate oxygen measurement is impossible by temperature change. Wait until the
instrumentmaybeusedtotheambienttemp(About10min)andstartmeasurementafter
havingmadezerocalibrationagain.
・Donotuseforafterpurposethenthemeasurementofcombustiblegas,vaporandtoxic
gasmixchares.
*Note
WhenmeasurethehighdensitygasthanLEL,thehung‑upphenomenon.
① WhenmeasureHCbyvol%,domeasureitasitis.
② Whenmakezeroadjustment,removethesamplingbaseandsupplyfreshairsuckedinto
instrument.
③ Whenmeasureby%LEL,makeaircleaningofsamplinghoseandstarttomeasureafter
readinggoesdowntozero.