PT0E‑0730
12
3−6
Operationalprocedure
Approachthesamplinghosetothesamplingsiteformeasurementandleavesamplinggas
sucked.
Takecarethatanytwistonthesamplinghosecannotbemade.
Stopsamplinghoseatthepointofmeasurement.
Fromthehoselengthconsiderthedelaytimeandreadoutthereadingafterthereading
getsstable.
Danger
Oxygendepletionairetcmaybeexhaustedfromgasoutlet.Donotbreatheinthere
absolutely.
Highdensitygasover(LEL)maybeexhausted.Donotapproachfirethereabsolutely.
Caution
Whensuckaircontaininggas,doitafterremovingwatervapours.
Whensuckhightempgas,doitafterbeingusedtobeambienttemp.
Donothavewaterandoilsuckedwrongly,theinternalpumpandsensormaybetroubled.
Whenmeasure,checktheroaringsoundofpumpandflowmonitor.
Itisimpossibletomeasureatpumpworkstop.
Donotblockthepumpoutlet,thereadingshallshowalittlehigher.
Whenmeasureoxygenininertgasforalongtime,besuretomeasurelowerthan15%CO2
byallmeans.Whenexeedhigherdensitythan15%,shortenthetimeforthismeasurement
aslifeofoxygensensormaybeshortend.
Warning
Intheplacewhereenvironmenttempisremarkablychanged,theremaybethecasethat
accurate oxygen measurement is impossible by temperature change. Wait until the
instrumentmaybeusedtotheambienttemp(About10min)andstartmeasurementafter
havingmadezerocalibrationagain.
Donotuseforafterpurposethenthemeasurementofcombustiblegas,vaporandtoxic
gasmixchares.
Note
WhenmeasurethehighdensitygasthanLEL,thehung‑upphenomenon.
WhenmeasureHCbyvol%,domeasureitasitis.
Whenmakezeroadjustment,removethesamplingbaseandsupplyfreshairsuckedinto
instrument.
Whenmeasureby%LEL,makeaircleaningofsamplinghoseandstarttomeasureafter
readinggoesdowntozero.
All manuals and user guides at all-guides.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments RIKEN RX-516 and is the answer not in the manual?

RKI Instruments RIKEN RX-516 Specifications

General IconGeneral
BrandRKI Instruments
ModelRIKEN RX-516
CategoryGas Detectors
LanguageEnglish

Related product manuals