PT0E‑0730
27
Filterreplacement
Forfilter(cotton)insideoffilterwithflowmonitor,checkitbeforeuseandwhenitgets
dirty,replaceit.
1Turntheknurlingpartindepthforfiltertubewithflowmonitorandopentheinside.
2Removethecottonofitbytweezersetc.
3Putthecottonfilteruniformlyonevenbaseinsideoffiltertubeandreturnbackthe
filtertubeinreverseprocedure.
Sensorreplacement
Whenthefollowingphenomenontookplace,itwillbethelifeendofsensor.
(1)Itcannotcalibrate.
(2)Readingcannotgobackifmakeaircalibration.
(3)Thereadingisunstable.
Whenthesensorreplacementisneeded,contactRIKENKEIKIornearestagent.
Dailycheckandregularcheck
1Daily
Isthereanydamageforswitches,lamps,displayandpanel?
Pumpsuctionperformancecheck(Doesthepumpmovingsoundcorrectly?Cananyabnormal
soundbeheard?)
Performancecheckofflowmonitor(CheckthatthepositionofflowiscorrectatpumpON
andOFF.
Batteryvoltagecheck
2Regularmaintenancecheck
Theregularmaintenanceisrecommendedtoreceivefromthequalifiedserviceagentoncein
ayear.
Caution
Donotputcottonfiltertoomuchinside.
Whensuctionflowistoomuch,itwillbedifficulttomonitoraccurately.
Note
Thecottonfilterreplacementshouldbemakeonceinamonth.
Knurlingcap
All manuals and user guides at all-guides.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments RIKEN RX-516 and is the answer not in the manual?

RKI Instruments RIKEN RX-516 Specifications

General IconGeneral
BrandRKI Instruments
ModelRIKEN RX-516
CategoryGas Detectors
LanguageEnglish

Related product manuals