PT0E‑0730
15
3−10.Practicalapplication
(1)Tankinsideofcrudeoiltanker
Byuseofsamplinghosewithplummet,itisavailabletomonitorthegasdensityinthetank.
《Sampleincrudeoiltanker》
《Useofsamplinghosewithplummet》
①Mountthesamplinghosetotheinstrument.
②Putintothetankfromtheplummetofhose.
③Samplinghoseshowsthemarkatevery5mandstopitatthepointdesiredtostopbychecking
thedistance.
④Checkthattheflowmonitorworksnormally.
⑤Whenthereadinggetsstableafter3minuteshavepassed,readoutthevalue.
⑥Whenchangethesamplingpoint,stopatthesamplingpointandinthesamemanner,read
outthereadingafterover3minuteshavepassed.
《HCreadout》
HCinthemeasuringgasstandsforiso‑butane.Butthecrudeoiltobemeasuredisamixtures.
Itisnotalwaystocorrespondtothegasdensity.Here,byexerptingthecrudeoilvapor
tobeconsidered,whenreadoutthevaluerepresentingnormalhexane,itwillbenearlyvalue
inreality.Incaseofnormalhexane,fromthevaporcondition,thisgascannotbevaporiged
tohighdensityrange(vol%).ThentheHCgasinthisdescriptionshouldbeiso‑butane
gas.
《H2Shighdensity》
HighdensityH2Smayeffectothersensorsfortrouble.So,intheatmosphereofhighdensity
H2Stobegenerated,RX‑517isrecommended.
Caution
WhenmeasureO2ininsertgasforalongtime,besuretouseCO2intheatmosphereof
below15%.Whenusehigherdensitythan15%,makethemeasuringtimeasshortaspossible.
COMBINATION GAS
DETECTOR
GAS SAMPRING PIPE
GAS
SAMPRING
PIPE
AIR VENT HEAD
WBT
COMBINATION GAS
DETECTOR
GAS SAMPRING PIPE
GAS
SAMPRING
PIPE
AIR VENT HEAD
WBT