PT0E‑0730
30
Troubleshooting
Thistroubleshootingdoesnotmentionallthetroublecauses.Thefollowingismentioned
brieflybypickingupthepointtohelpsurveyoftroublecauseinhighfrequency.
Phenomenon Cause Treatment
SYSTEMFAIL Excessnoiseeffectsetc. Re‑start up and check status. Request to
manufacture.
Sensortrouble Checkconnectstatusandcontactmanufacture.FAILSENSOR
・D i d excessshocketcbydropping,
crushing.
Checkconnectstatusandcontactmanufacture.
Water,oilaresoaked. Checkwhetheranydamage,suctionmarksofwater,
oiletcshallbefoundonsamplinghoseandfilter
withflowmonitor.
Fillingbyfilterdust. Replacethefilterwithnew.
FAILFLOW
Substanceisclogged. Removesubstance.
LOWBATTERY
(REPLACEBATT.)
Lowbatteryofdrybattery. Makepoweroffandreplacebatteryinnon‑hazardous
area.(See6‑1.BATT.REPLACEMENT.)
Freshairisnotsuckednormally. Makezeroadjustmentagainbyhavingfreshair
suckedintoinstrument.See3‑5.ZEROADJUSTMENT)
ZEROCALFAIL
(FAILZERO)
Sensortrouble. Makesensorreplacement.
・C a l gasisnotsuppliedcorrectly Make gas calibration again by supplying cal
gas.(See6‑2.GASCALIBRATION.)
SPANCALFAIL
(FAILSPAN
Sensortrouble. Makesensorreplacement.
Nopoweron. Putbatterycorrectly.
Batteryisexhausted. Replacebatteries.
Polarityiswrong. Correctbatterypolarity.
PowercannotbeON
PressingPOWERSWistooshort. HoldandpressPOWERSWfor3sec.
Pumpisworking
PowercannotbeON
Allthe batteries are put in
reversely.
Put batteries correctly according to battery
replacement(See6‑1.BATTERYREPLACEMENT.)
Pumplifeendsortroubled. Requestmanufacturetorepair.
Not measured and pump stop
working.
Re‑startpumpoperation.PressPUMPSWuntilLCD
willdisplayPUMPON.
Pumpdoesnotwork
Pumpconnectionfails. CheckconnectionconditionandcontactRIKENKEIKI
forrepair.
Gasisnotsucked Filterorhoseetcareremoved
orclogged.
Checkconnectionstatusoffilter,hoseandfilth
ortwistanduseincorrectcondition.
Zeroblinks Zeroisbelowminus. Check that gas is free around and make zero
adjustment.
ForCO2,makeitaftermountingCO2scrubber.
100.0vol%blinks
(HConly)
Zeroandspanaredeviatedand
itshallbetheoutputof100
vol%
Repeatagaintheprocedureofzeroadjustmentand
gascalibration.
All manuals and user guides at all-guides.com
all-guides.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments RIKEN RX-516 and is the answer not in the manual?

RKI Instruments RIKEN RX-516 Specifications

General IconGeneral
BrandRKI Instruments
ModelRIKEN RX-516
CategoryGas Detectors
LanguageEnglish

Related product manuals