PT0E‑0730
32
10.Specifications
Refertothespecificationsforseparatecatalogoverleaf.
Model RX-516(HC/O2/H2S0-100.0ppm),RX-517(HC/O2/H2S0-100.0ppm&0-1000ppm)
Gasdetected Iso-butane(HC) Oxygen(O2) Hydrogensulfide(H2S)
Detection
principle
NDIR(Non-Dispersive
infrared)
Galvaniccell Electro-chemicalcell
Detectionrange
(digit)
0-100%LEL(0.5%LEL)
2-100vol%(0.5vol%)
0-25.0vol%
(0.1vol%)
<LOW>
0-100.0ppm
(0.5ppm)
<HIGH>
0-1,000ppm(1ppm)
OperatingTemp.&
Humidity
-10 〜 +50℃,below90%RH(non-condensing)
Safetyrating ExiadIIBT3X(AlkalinebatteriesSizeConly),KIMMapproved,MEDapproved
Powersource Alkalinebatteries(SizeC)4pcs
Continuous
operation
Approx.30hours(noalarmorlight,at25℃)
Display DigitalLCD(withbacklight)
Alarmtypes
Gasalarm:Standard……OFF
(Optional……1
st
and2
nd
useradjustable,self-resetting)
Troublealarms(Selfdiagnosis):Sensordisconnection,lowbattery,circuittrouble
Andcalibrationerror
Dimension&
Weight
Approx.200(W)x80(H)x142(D)mm
Approx.1.8kg(includingbatteries)
Approx.2.0kg
(includingbatteries)
Specialfunctions
Peakhold,Datalogging
Highconcentrationhydrogensulfiderangewithmanualswitch(RX-517)
Standard
accessories
①Carryingcasewithshoulderstrap, ②Filtertubewithflowmonitorandjunctiontube,
③Gas sampling hose, ④Gas sampling probe, ⑤Alkaline batteries (4pcs), ⑥Instruction
manual