Troubledisplay
1Lowflowalarm
Whenlowflowtroubletriggers,LOWFLOWisdisplayedandpumpstopsworking.
Buzzersoundscontinuously.
Aftercheckingthecause,resetlowflowalarmbyESCkey.Thepumpwillre‑start
andreturntotheoperationdisplay.
2Lowbatteryalarm
Whenthepowerofbatteryisdropped,REPLACEBATT.Willbedisplayed.
Buzzerwillsoundcontinuously.
AfterdisplayingREPLACEBATT.,itdoesnotacceptotherthanPowerOFF.
Afterpoweroff,replacethebatterywithnewone.(Seeitfor6‑1.Batteryreplacement).
REPLACEBATT.
Lowbattery
POWEROFFoperationavailableonlybyPOWERkey.
LOWFLOW
Pum
p
OFF
ESCkey
Pum
p
ON
Caution
Whenlowfloattroubletookplace,besuretocheckthecausetomakelowflow.
Especiallywhensuckwater,itwillbethecausefortrouble.Aftercheckingthecondition
ofprobefiltertube,re‑startthepumpbyESCkey.(WhenpressESCandre‑startinthe
waterremainingcondition,therewillbethecaseofwaterintrusionintoinstrument.
Forlowflowtrouble,itshallbecautionedthatnoerrordisplaywillbethesettingmode.
All manuals and user guides at all-guides.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments RIKEN RX-516 and is the answer not in the manual?

RKI Instruments RIKEN RX-516 Specifications

General IconGeneral
BrandRKI Instruments
ModelRIKEN RX-516
CategoryGas Detectors
LanguageEnglish

Related product manuals