《Calgasmake‑upbyH2SgascheckerCK‑82》
① Takeout1pce.ampoulefromaccessoryampoulepack.
② Checkthattherubbercapsarefixedinthecoverofcontainer.
③ Openthecoverofcontaineranddeflatetheairbysqueezingthepressureadjbag.
④ Putuptheampoulecrusherknowandputtheampouleintoampoulecrusherinsideofthe
covershownbelow.Whentheampouleisputinreversely,theampoulemaybecrushed.
⑤ Aftertheampouleisputin,closethecover.
⑥ Checkthattherubbercapsareputonthetopcover.
⑦ Aftercheckingthatthecoverisclosed,putintheampoulecrusherknobandcrushthe
insideampoule.
⑧ Gripthecontainerwithhandsandshakeittoup&downleft&rightslightlysothat
H2Scannotremainintheampoulecrushed.
⑨ About21ppmH2Sissupposedtobeproducedinthecontainer.
Caution
・21ppmH2SgasdensitybyCK‑82isthetheoreticalvalueinthecalculation.Actually,
theerrorof1‑2ppmcouldbeexpected.WhendesiredtocheckH2Sinthecontainer,
checkitbydetectortube.WhenusegascheckerCK‑82,seetheseparatemanualfor
modelCK‑82together.
Ampoule
Ampoulecrusherknob
Crushtheampoulebypressingtheampoulecrusherknob