PT0E‑0730
25
Handlingofsamplingbagandcannedgas》
Samplethecalgasintosamplingbagfromcannedgas.
Samplingbagwithcalgasinshallbeusedbyputtingtogasinsideatgascalibration.
Cannedgas
Samplingbag
Danger
Donotmakegascalibrationbyuseoflightergas.Bytheingredientinthelighter
gas,thesensorfunctionmaydeteriorate.
Caution
Donotsupplythecalibrationgasbypluggingintheinletofcannedgasdirectly.
Itmayleadtotheinternaldamageofinstrument.
Danger
Compressedcylindergasisusedforgascalibrationandtakecareathandling.
Atcalibrationbycombustiblegas,donotapproachittothefire.
Exhausttheremaininggasinthesamplinggasappropriately.
Caution
Beforegascalibration,besuretomakezeroadjustment.