씰 Wipe the measurement optics components with a soft lint-free
cloth and/or a cotton swab. The cloth/cotton swab may be
slightly moistened with water or 70% alcohol. Make sure that no
liquid enters the meter itself. Avoid scratching the measurement
optics, as this will impair the measuring function.
씰 When all of the components are thoroughly dry, you may slide
the test strip guide back onto the meter. Ensure it clicks back
into place.
Then perform a quality control test (see Chapter 6, “Checking
Accu-Chek Active”).