87
The meter performs a display test lasting 2 seconds.
Check that all the segments making up the numerals (“
888
or “
88.8
”) are properly displayed. If a segment is missing,
test results may be displayed inaccurately (e.g. through
9
being
confused with
3
). If this happens, call your customer support
and service centre.
The current code number then appears in the display. Is this
the number printed on the test strip container? If not, check
that you really did insert the coding chip from the new pack.
If “code” is flashing and you see three horizontal bars (
---
)
instead of a number, you have not inserted the coding chip.
You can still do this now (while the display is flashing).
Check that the correct date and time are displayed.