Control Testing and Quality Control
99
7 Apply gentle pressure to the reservoir to
transfer the entire contents of the dropper
to the bottle. Make sure that the dropper
does not come into contact with the dried
control plasma.
8 Close the bottle again.
9 Make sure the dropper is at hand for the
next steps in the liquid quality control test.
10 Swirl the bottle using a circular motion to
completely dissolve all of the control
plasma inside. Do not shake the bottle
or turn it on its side. Doing so can cause
components in the control plasma to stick
to the sides of the bottle. Let the bottle sit
undisturbed for 1 minute. Please refer to
the control solution package insert.
The control solution is now ready to be applied
to the test strip.
The control solutions may be reconstituted
(mixed) after removal from the refrigerator.
The resulting solution may be used up to 30
minutes after reconstitution.