Supplement for Observed Test Sequence
140
Using the OTS function
A request for an Observed Test Sequence comes from the DMS. The presence of the icon
on the Patient Test button indicates a pending OTS request.
Observer:
1 Touch Patient Test.
In the Pat. Test -OTS menu, the Patient Test
button is grayed out (disabled) until the
observer has logged in.
2 Touch Observer Login.
3 Wait until the observer list is displayed.
4 Select your observer ID by touching the
corresponding button.
Only operators with OTS observer rights are
listed in the Observer Login list.
5 Enter the (optional) password.
6 After you enter your password, touch
to log in. The Pat. Test - OTS menu is dis-
played again. The Patient Test button is
now active.
7 Hand the meter to the operator who can
now perform the patient test under
supervision.
Main Menu 09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Logout
Pat. Test - OTS 09:15 am
04/19/2012
Observer Login
Patient Test
OTS enabled
Observer 09:15 am
04/19/2012
Ann Casey
Alan Smithee
Pat. Test - OTS 09:15 am
04/19/2012
Observer Login
Patient Test
OTS enabled