Setting up a Run for Purification of Genomic DNA from Whole Blood Samples
35
B
B
Loading the Stage
In the shaking section of the bottle rack, pull back the clip of the rearmost bottle holder (position 1) and
insert the first MGP bottle.
Press the bottle firmly into the holder until the clip slips over the bottle shoulder.
Insert a second MGP bottle in position 2 of the bottle rack according to the Stage sub-tab in the same
way.
Ensure that the clips fix the bottle rims completely and that the barcode labels of all bottles face left so
that they can be read by the barcode reader during the resource check.
Especially if you use the same MGP bottles during subsequent runs, the bottles can rotate during
the shaking process, so that the barcode may no longer be readable for the barcode reader.
Remove the cap from the prepared internal control tube and insert it in position 9 of the bottle rack.
Insert the bottle rack back into the stage.
Press the bottle rack firmly into the stage until its outline is displayed in blue on the Stage sub-tab.