MagNA Pure 96 System, Customer Training Guide V1.0
36
Setting up a Run for Purification of Genomic DNA from Whole Blood Samples
B
B
Loading the Stage
To load the reagent racks
The reagent racks hold the processing cartridges and the reagent trays needed for the defined order. The
processing cartridges and reagent trays are hardware-coded to guide correct positioning:
The Stage sub-tab displays the reagent racks according to the selected purification protocol with the
Relevant number and position of processing cartridges
Number and label of the reagent trays
Position of the reagent trays in the reagent racks
Processing cartridge The bottom left frame has a rectangular cut that fits onto an elevation in the
relevant rack position.
Reagent tray A groove in the bottom matches a tongue at the correct position in the re-
agent rack.
Pull the left reagent rack all the way out of the stage.
Place a processing cartridge in position 1 of the left reagent rack.
Ensure that the barcode label faces left so that it can be read by the barcode reader during the resource
check.
Insert the left reagent rack back into the stage and press it in until its outline is displayed in blue on the
Stage sub-tab.
Pull the right reagent rack all the way out of the stage.