R&S CMU MODULE Replacement
1100.4903.82 3.63 E-5
Installing the new sandwich module and completing the instrument
Replace by new module in reverse order.
Caution!
Take care to use the torque screwdriver tool
to fit the screws with the right
torque.
Take care to insert the respective module with the correct polarity (see chapter
5, drawing 1150.0301.01 sheet 2).
Replace the covers and fasten with screws.
Insert the Option CDMA2000 SIGNALING UNIT into the instrument in the reverse order and connect
all the cables on the top.
Complete the instrument without causing damage to the cables.
Connect the instrument to the mains and switch on.
Automatic module data adjustment
During startup observe the display.
When three BEEPs can be heard, press the MENU SELECT key.
The version manager is started (see also chapter 4, Firmware update). The display includes the
menu item FIRMWARE UPDATE AFTER BOARD CHANGE.
Press the softkey to the left of the above mentioned menu item.
The automatic module data adjustment is performed under software control, firmware updates for
microprocessors and programmable devices being performed as well. This may take a few minutes.
After the adjustment has been terminated, the operating software starts automatically and the R&S
CMU is ready for use and complies with the specifications.
available in CMU-Service Kit CMU-Z3 (1100.7690.02)

Table of Contents

Other manuals for Rohde & Schwarz CMU 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz CMU 200 and is the answer not in the manual?

Rohde & Schwarz CMU 200 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelCMU 200
CategoryTest Equipment
LanguageEnglish

Related product manuals