R&S CMU Manual Test Procedure
1100.4903.82 1.11 E-5
TX Frequency Accuracy
Preparation: Test setup TX5, but R&S CMU not synchronized with frequency counter/analyzer
and no external trigger.
R&S CMU connector RF3 OUT.
Control: Set R&S CMU to desired frequency, level 0 dBm.
Test: Determine frequency deviation from nominal frequency.
TX Frequency Settling Time
Preparation: Test setup TX5, in addition trigger cable from R&S CMU (D-sub connector AUX3,
pin2) to analyzer.
R&S CMU connector RF3 OUT.
R&S CMU: Ramping off, hopping on, F1 = start freq., F2 = stop freq.
Analyzer: Sweep time 1 ms, Center = stop frequency, FM demodulation,
real time off, BW 50 kHz, 1 kHz/Div, external trigger, Slope
negative
Control: Set R&S CMU to desired frequencies and hopping, level 0 dBm.
Test: Time from trigger point when the specified offset (< 1 kHz) from the stop
frequency is reached.
TX Level Error
Preparation: Test setup TX1 to TX4 (depending on level range).
Control: Set R&S CMU to desired connector, frequency and level (RF Analyzer must
be OFF).
Test: Measure the TX level of the R&S CMU.
Note: The given frequencies and levels are suggested values. Of course, it is also possible to use
other values for the measurement.
VSWR
Preparation:
Connect (scalar) network analyzer to RF1, RF2, RF3 OUT, RF4 IN one after
the other.
Cable losses must be corrected.
Control:
R&S CMU: Switch generator on and set level to minimum (–130 dBm or
–90 dBm), switch RF wideband analyzer on (RF1/RF2/RF4 IN).
Test: Measure VSWR at 10 MHz, 900 MHz, 1800 MHz, 2700 MHz.

Table of Contents

Other manuals for Rohde & Schwarz CMU 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz CMU 200 and is the answer not in the manual?

Rohde & Schwarz CMU 200 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelCMU 200
CategoryTest Equipment
LanguageEnglish

Related product manuals