R&S CMU Test Report
1100.4903.82 1.43 E-5
Item
No.
Description
Measurement
to section
Min. Actual Max.
Measurement
uncertainty
Unit
K23 TX GSM phase error
at 1990 MHz, rms
–1 +1 °
K23 TX GSM frequency error
at 1990 MHz
–15 +15 Hz
K24 TX GSM phase error
at 869 MHz, peak
–4 +4 °
K24 TX GSM phase error
at 869 MHz, rms
–1 +1 °
K24 TX GSM frequency error
at 869 MHz
–15 +15 Hz
K24 TX GSM phase error
at 894 MHz, peak
–4 +4 °
K24 TX GSM phase error
at 894 MHz, rms
–1 +1 °
K24 TX GSM frequency error
at 894 MHz
–15 +15 Hz
Options for R&S CMU200: R&S CMU-K20, K21, K22, K23, K24, RX Analyzer GSM Demodulation
Input RF2, GSM Non Signaling
Training Sequence GSM0, Trigger Source IF Power, Trigger Level Medium
Item
No.
Description
Measurement
to section
Min. Actual Max.
Measurement
uncertainty
Unit
K20 GSM phase error RX
at 450 MHz, peak, level +5 dBm
RX
Demodulation
–2 +2 °
K20 RX GSM phase error
at 450 MHz, rms, level +5 dBm
–0.6 +0.6 °
K20 RX GSM frequency error
at 450 MHz, level +5 dBm
–10 +10 Hz
K20 RX GSM phase error
at 486 MHz, peak, level –14 dBm
–2 +2 °
K20 RX GSM phase error
at 486 MHz, rms, level –14 dBm
–0.6 +0.6 °
K20 RX GSM frequency error
at 486 MHz, level –14 dBm
–10 +10 Hz
K21 RX GSM phase error
at 876 MHz, peak, level +5 dBm
RX
Demodulation
–2 +2 °
K21 RX GSM phase error
at 876 MHz, rms, level +5 dBm
–0.6 +0.6 °
K21 RX GSM frequency error
at 876 MHz, level +5 dBm
–10 +10 Hz
K21 RX GSM phase error
at 915 MHz, peak, level –14 dBm
–2 +2 °
K21 RX GSM phase error
at 915 MHz, rms, level –14 dBm
–0.6 +0.6 °