Instrument Design and Function Description R&S CMU
1100.4903.82 3.12 E-5
The transmitter side is provided with an extra LO1TX (Local Oscillator)
w
ith large tuning range and very fine frequency resolution used for
setting the desired transmitter frequency, an LO2 fixed-frequency
oscillator shared with the receiver side and an extra LO3TX with small
tuning range. All LOs are synchronized by the MMCX connector
110.8MHz at the bottom of the module with the reference frequency
from the REFERENCE BOARD.
Besides, an MMCX connector at the top of the RXTX BOARD1 is
designed as input with the second transmitter IF (RESIF2TX) which is
activated via software switches.
Receiver functions
On the receiver side (RX), the RF signal is fed in at the SMA connector
RFRX1 (bottom of module) of the RXTX BOARD1 which is provided by
the RF FRONTEND.
On the RXTX BOARD1, this signal is taken via a switchable attenuator
with fine stepping and large attenuation range in order to match the
level according to the level applied.
Subsequently, this signal is converted to an IF frequency of 10.7 MHz
by means of triple signal conversion (IF1, IF2, IF3) and several filter
and amplifier stages and provided at the MMCX connector IF3RX1 (top
of module) for the DIGITAL BOARD.
This procedure is necessary to achieve a high image-frequency
rejection and a high dynamic range with a simultaneously high
intermodulation suppression.
For the receiver side, an extra LO1RX (local oscillator) is provided with
a large tuning range and a very fine frequency resolution used for
setting the receive frequency, an LO2 fixed-frequency oscillator shared
with the transmitter and an extra LO3RX with a very small tuning
range.
All LOs are synchronized by the MMCX connector 110.8 MHz at the
bottom of the module with the reference frequency from the
REFERENCE BOARD.
A further MMCX connector RESIF3RX (top of module) provides a
buffered RX IF signal of 10.7 MHz which is taken to the rear of the
R&S CMU via a coax cable. This signal output can be activated via a
software switch.
Besides, an MMCX connector at the top of the RXTX BOARD1 is
designed as output with the second receiver IF (RESIF2RX) which can
be activated via software switches.
Correction processor
The RXTX BOARD1 contains an extra correction processor with large
flash PROM.
It controls all the static and dynamic settings on the RXTX BOARD1
and, as a special feature, also the attenuator pads and amplifiers of the
RX and TX attenuator on the RF FRONTEND.
Besides, the correction processor permits to read out the individual
module error data from the EEPROMS of the respective modules in a
so-called correction procedure (automatic module data adjustment)
and calculate the deviations for all possible signal paths. These
deviations are stored as correction values in the flash PROM. The
correction processor then sets the desired level settings, corrected by
the correction values, in the flash PROM so that frequency, linearity
and temperature responses of the modules are compensated for.
This ensures the excellent level accuracy of the R&S CMU which is
essential for most measurements.
Extension
The R&S CMU 200 can be extended by a RXTX BOARD2 and thus
provide a second complete transmit and receive channel.

Table of Contents

Other manuals for Rohde & Schwarz CMU 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz CMU 200 and is the answer not in the manual?

Rohde & Schwarz CMU 200 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelCMU 200
CategoryTest Equipment
LanguageEnglish

Related product manuals