EasyManua.ls Logo

Rohde & Schwarz FSP3 - Diplexer Replacement Procedures

Rohde & Schwarz FSP3
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Module Replacement R&S FSP
1164.4556.82 3.40 E-1
Replacing Diplexer A162; R&S FSP7
(See chapter 5, spare parts list, item 2030, and drawing 1130.2544.)
The diplexer is installed on the 8 GHz converter unit (1130.2544)..
Removing Diplexer on MW Converter 1130.2544
Remove 8 GHz converter unit (130).
Note: See description "Replacing 8 GHz Converter Unit A160"
Disconnect cable W1 from diplexer (2030).
Undo 4 screws with washers (2035) from the bottom of the PCB.
Remove diplexer (2006) perpendicularly to the PCB.
Note: The diplexer is still connected to the PCB with contact pins.
Installing Diplexer on MW Converter 1130.2544
Plug in new diplexer (2030) perpendicularly to the PCB.
Note: The diplexer is connected to the PCB with contact pins.
Refasten 4 screws with washers (2035) to the bottom of the PCB.
Fit cable W1 to diplexer (2030).
Completing the Instrument and Frequency Response Correction
Insert 8 GHz converter unit (130) back into the instrument.
Note: See description "Replacing MW Converter Unit A160".
Execute frequency response correction software R&S FSP-FRQ.EXE.
The software is located on the Gloris server.
- Test setup according to menu item "Schematic" (connect external reference).
- Check IEC/IEEE addresses and configuration of instruments.
- Press Autoselect button and perform the measurement with Run.
- The frequency correction of the YIG filter is performed under item YIG.
- After calibration, the frequency response is corrected over the entire frequency range.

Table of Contents