EasyManua.ls Logo

Rohde & Schwarz FSP3 - Checking Frequency Modulation

Rohde & Schwarz FSP3
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Test R&S FSP
1164.4556.82 1.26 E-1
Checking Frequency Modulation
Test equipment:
- Arbitrary waveform generator (refer to "Measurement Equipment",
item 16)
- Spectrum analyzer (refer to "Measurement Equipment", item 15)
- 2 voltmeters (refer to "Measurement Equipment", item 17)
- 2 BNC T pieces
Test setup with R&S ADS: Apply the I and Q outputs of ADS to the I and Q inputs of the
analyzer. Check the ADS voltages in parallel via T pieces.
Connect spectrum analyzer to the TG output.
Test setup with 33220A: Apply the outputs of 33220A to the I and Q inputs of the analyzer.
Check the 33220A-
voltages in parallel via T pieces.
Synchronize the two 33220A via Option 001.
Connect spectrum analyzer to the TG output.
Generator settings:
AC voltage: V(pp) = 100 mV ± 10 mV
Frequency: 100 kHz
Spectrum analyzer settings:
- [ FREQUENCY : 1 GHz ]
- [ SPAN : 100 MHz ]
- [ BW : 1 MHz ]
- [ REF : REF LEVEL : 0 dBm ]
R&S FSP settings:
- [ FREQUENCY : 1 GHz ]
- [ SPAN : 0 MHz ]
- [ REF : REF LEVEL : 0 dBm ]
- [ NETWORK : SOURCE POWER 0 dBm ]
- [ MODULATION EXT FM ]
Measurement: Determination of the peak spacing = 2 x DEVIATION