R&S FSP Performance Test
1164.4556.82 1.11 E-1
Checking Level Accuracy and Frequency Response
Test equipment:
- Signal generator:
R&S FSP 3: Refer to "Measurement Equipment", item 3.
R&S FSP 7/13/30:Refer to "Measurement Equipment", items 3 & 4.
Frequency range R&S FSP 3: 1 MHz to 3 GHz
R&S FSP 7: 1 MHz to 7 GHz
R&S FSP 13: 1 MHz to 13.6 GHz
R&S FSP 30: 1 MHz to 30 GHz
R&S FSP 40: 1 MHz to 40 GHz
Maximum level 0 dBm
- Power meter (refer to "Measurement Equipment", item 8)
- Power sensor:
R&S FSP 3: Refer to "Measurement Equipment", item 9.
R&S FSP 7/13/30: Refer to "Measurement Equipment",
items 9 & 10.
Frequency range R&S FSP 3: 1 MHz to 3 GHz
R&S FSP 7: 1 MHz to 7 GHz
R&S FSP 13: 1 MHz to 13.6 GHz
R&S FSP 30: 1 MHz to 30 GHz
R&S R&S FSP 40: 1 MHz to 40 GHz
Maximum power P
max
100 UW
RSS referenced to indicated power
1 MHz to 1 GHz 1.5 %
1 GHz to 7 GHz 2 %
7 GHz to 13.6 GHz 3.5 %
13.6 GHz to 40 GHz 4 %
Impedance Z = 50
- Power splitter (refer to "Measurement Equipment", item 6)
Frequency range R&S FSP 3: 1 MHz to 3 GHz
R&S FSP 7: 1 MHz to 7 GHz
R&S FSP 13: 1 MHz to 13.6 GHz
R&S FSP 30: 1 MHz to 30 GHz
R&S FSP 40: 1 MHz to 40 GHz
Level imbalance
1
) 1 MHz to 1 GHz 0.1 dB
1 GHz to 7 GHz 0.2 dB
7 GHz to 13.6 GHz 0.3 dB
13 GHz to 40 GHz 0.4 dB
1
) If a power splitter with higher level imbalance is used, correction
of the measured frequency response is recommended.