EasyManua.ls Logo

Rohde & Schwarz FSP3 - DC;DC Converter Replacement

Rohde & Schwarz FSP3
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Module Replacement R&S FSP
1164.4556.82 3.66 E-1
Replacing the DC/DC Converter Option R&S FSP-B30)
(See chapter 5, spare parts list, item 3000.)
The DC/DC converter (3000) is mounted on the rear panel of the instrument.
See also DC Power Supply R&S FSP-B30 operating manual.
CAUTION
The DC/DC converter in an unmounted state is a module in "open frame" design.
Therefore make sure that it is mounted only by skilled electricians (see operating
manual, safety instructions, item 9).
Make sure that the DC/DC converter is separated from the supply source.
Removing the DC/DC Converter
Unplug the supply cable on the DC/DC converter.
Unscrew the three captive screws on the converter.
Pull off the converter to the rear.
Installing the new DC/DC Converter
Plug the DC/DC converter (3000) onto the rear-panel power plug.
Note: The power switch is automatically actuated when the DC/DC converter is connected.
Fasten the converter to the rear panel with three fixing screws.
Note:
Rear-panel foot with Options R&S FSP-B30 (DC/DC converter) and R&S FSP-B1 (housing with
shock protection):
If Option R&S FSP-B1 is present, the right rear-panel foot (3020) is modified in order to ensure sufficient
ventilation of the power pack.
Fig. 3-1 Removing and installing the DC/DC converter, R&S FSP-B30