Temperature Adjustment and Test of the YIG Filter 
 A digital temperature adjustment is to be performed for the following MW converters: 
- MW converter unit 13/30: 1093.8249  Mod: XX   Rev:  10  SubRev:  01 
- All other MW converters should undergo analog adjustment. 
 Analog adjustment:  
- Connect the RF output of the signal generator to the RF input of the R&S FSP. 
- Activate the YIG filter sweep with the service function: 
  Password    30473035 
  Service funct.  2.16.10.1 
- Span on R&S FSP30: 200 MHz 
- Set the frequency of the signal generator and the center frequency to 15 GHz. 
Nominal value: Temperature drift (0 to 50°C) < 8 MHz 
- Adjust temperature drift with jumpers X21, X22, X23, trimming values R403, R119, R402 and 
potentiometer R232.
 
 Digital adjustment: 
- Place R&S FSP30 into a climatic chamber for 2 hours at 50°C. 
- Execute frequency response correction R&S FSP-FRQ.EXE. 
The software is located on the Gloris server. 
- Test setup according to menu item "Schematic" (connect external reference). 
- Check IEC/IEEE addresses and configuration of instruments. 
- Press the TK adjustment button. 
- Place R&S FSP30 into a climatic chamber for 2 hours at 0°C in the switched-off state. 
- Switch on the instrument and repeat the coldness adjustment. 
 
Adjustment of Module Data 
 
 Create a backup of the EEPROM data (see chapter 2, section "Adjustment of Module Data").