EasyManua.ls Logo

Rohde & Schwarz FSWP - Self-Alignment and Test Setup Considerations

Rohde & Schwarz FSWP
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preparing for Use
R&S
®
FSWP
30Getting Started 1177.5756.02 ─ 13
3.13 Performing a Self-Alignment
When strong temperature changes occur in the environment of the R&S FSWP,
or after updating the firmware, you have to perform a self-alignment to align the
data to a reference source.
During self-alignment, do not connect a signal to the RF input connector. Running
a self-alignment with a signal connected to the RF input can lead to false mea-
surement results.
Performing a self-alignment
Before performing this functional test, make sure that the instrument has reached
its operating temperature (for details, refer to the data sheet).
A message in the status bar ( "Instrument warming up..." ) indicates that the oper-
ating temperature has not yet been reached.
Depending on the installation settings, an automatic self-alignment is performed
each time the instrument is switched on. A dialog is displayed indicating how
much warm-up time is still required before self-alignment can be performed.
1. Press the [Setup] key.
2. Press the "Alignment" softkey.
3. Select the "Start Self Alignment" button in the "Alignment" dialog box.
Once the system correction values have been calculated successfully, a mes-
sage is displayed.
To display the alignment results again later
Press the [SETUP] key.
Press the "Alignment" softkey.
3.14 Considerations for Test Setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
Considerations for Test Setup

Table of Contents

Related product manuals