EasyManua.ls Logo

Rohde & Schwarz R&S NRP-Z91 - Commands as per IEEE 488.2

Rohde & Schwarz R&S NRP-Z91
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands R&S NRP-Z91
1168.8579.12 6.2 E-3
Commands as per IEEE 488.2
The sensor supports a subset of the possible setting commands and queries (Common Commands and
Queries) in line with IEEE 488.2.
*IDN? Identification Query
*IDN? returns a string with information on the sensor's identity (device identification code). In addition,
the version number of the installed firmware is indicated. The string for a sensor of type R&S NRP-Z51
has the following structure:
ROHDE&SCHWARZ,NRP-Z51,<serial number>,<firmware version>
<serial number>: Serial number in ASCII
<firmware version>: Firmware version number in ASCII
*RST Reset
*RST sets the sensor to the default state, i.e. the default settings for all test parameters are loaded.
*TRG Trigger
*TRG triggers a measurement. For this purpose, the sensor is in the WAIT_FOR_TRIGGER state and
the source for the trigger event is set to BUS (TRIGger:SOURce BUS).
*TST? Self Test Query
*TST? starts a selftest and returns 0 (no error found) or 1 (an error has occurred). The selftest
comprises the following functions:
RAM test
Operating voltages
Temperature measurement
Calibration data set
Noise
Zero-point offsets.

Table of Contents

Related product manuals