EasyManua.ls Logo

Rohde & Schwarz R&S NRP-Z91 - Calibration:zero:auto[?] off | on | ONCE

Rohde & Schwarz R&S NRP-Z91
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control - Commands R&S NRP-Z91
1168.8579.12 6.4 E-3
CALibration:ZERO:AUTO[?] OFF | ON | ONCE
The commands CALibration:ZERO:AUTO ON and CALibration:ZERO:AUTO ONCE zeroes the three
measurement paths of the sensor. For this purpose, the test signal must be deactivated or the sensor
disconnected from the signal source. The sensor automatically detects the presence of any significant
power to be measured. This causes zeroing to be aborted and error message NRPERROR_CALZERO
to be output. The command CALibration:ZERO:AUTO OFF is ignored. Zeroing takes four seconds at a
minimum, but at least as long as the selected averaging filter needs for settling (only fixed-filter mode).
Repeat zeroing
during warm-up after switching on or connecting the instrument
after a substantial variation of the ambient temperature
after fastening the sensor to an RF connector at high temperature
after several hours of operation
when very low-power signals are to be measured, e.g. less than 10
dB above the
lower measurement limit.
For zeroing switch off the test signal and do not remove the sensor from the signal source.
Apart from keeping the thermal balance, this has the advantage that the noise
superimposed on the test signal (e.
g. from a broadband amplifier) can be detected on
zeroing and does not impair the measurement result.
The query always yields 1 (= OFF).
Default setting
After a power-on reset, the zero offsets determined during the last calibration are used until the first
zeroing. Therefore, very slight zero offsets are to be expected with a warmed up sensor. Initialization by
means of *RST or SYSTem:INITialize has no influence on the current zero offsets.

Table of Contents

Related product manuals