Remote Control - Commands  R&S NRP-Z91  
1168.8579.12 6.6  E-3 
Command  Parameter  Unit  Remarks 
:SMOothing:STATe[?] OFF | ON     
:RANGe[?] 0 to 2    
:RANGe:AUTO[?] OFF | ON     
:RANGe:CLEVel[?] –20.0 to 0.0  dB  
:SGAMma    
:CORRection:STATe[?] OFF | ON     
:MAGNitude[?] 0.0 to 1.0    
:PHASe[?] –360.0 to 360.0  degree  
SENSe:AVERage:COUNt[?]   1 to 65536 
SENSe:AVERage:COUNt sets the number of measured values that have to be averaged for forming 
the measurement result. The higher this averaging factor, the less the measured values fluctuate and 
the longer the measurement time lasts. The parameter is rounded off to the nearest power-of-two 
number. 
 
The query yields the set averaging factor. 
 
The averaging function must be activated with SENSe:AVERage:STATe ON
set averaging factor becomes effective.  
Default setting: 4
SENSe:AVERage:COUNt:AUTO[?]   OFF | ON | ONCE 
SENSe:AVERage:COUNt:AUTO activates (auto-averaging) or deactivates (fixed-filter mode) automatic 
determination of the averaging factor. If auto-averaging is activated, the averaging factor is continuously 
determined and set depending on the level of power and other parameters. 
 
SENSe:AVERage:COUNt:AUTO ON activates auto-averaging and SENSe:AVERage:COUNt:AUTO 
OFF deactivates it. On deactivation, the previous, automatically determined averaging factor is used in 
the fixed-filter mode. The SENSe:AVERage:COUNt:AUTO ONCE command ensures that a new 
averaging factor is determined by the filter automatic function under the current measurement 
conditions and used in the fixed-filter mode.  
 
The query yields 
• 1 for OFF, 
• 2 for ON.
Default setting: ON