R&S NRP-Z91 Remote Control - Commands
1168.8579.12 6.21 E-3
TRIGger:LEVel[?] x to y
TRIGger:LEVel sets the trigger threshold for internal triggering derived from the test signal (in W). This
setting is irrelevant to all other trigger sources. If an s-parameter device is active and/or if a component
with a global offset upstream from the sensor is considered, the currently effective trigger threshold as
well as a trigger threshold to be input are referenced to the appropriately shifted sensor interface. When
the s-parameter device and the offset correction are switched off, then the trigger threshold and its input
limits are adjusted as necessary.
The query yields the currently effective trigger threshold in W.
Lower limit x and upper limit y of the trigger threshold
SENSe:CORRection:OFFSet:STATe OFF:
x = <lower test limit in W> × 500
y = <upper test limit in W>
SENSe:CORRection:OFFSet:STATe ON:
x = <lower test limit in W> × 500 × 10
OFFSET / 10 dB
y = <upper test limit in W> × 10
OFFSET / 10 dB
<lower test limit in W>: The nominal lower test limit of the sensor or of the twoport/sensor
combination (with activated s-parameter correction)
<upper test limit in W>: The nominal upper test limit of the sensor or of the twoport/sensor
combination (with activated s-parameter correction)
The nominal test limits can be read out with the SYSTem:INFO? "MINPOWER" and SYSTem:INFO?
"MAXPOWER" command.
Default setting: 10 × x
TRIGger:SLOPe[?] POSitive | NEGative
TRIGger:SLOPe defines the edge of the trigger event with internal and external triggering in the
Continuous Average, Timeslot Average, and Trace modes. In this connection, positive means
increasing envelope power (with internal triggering) or increasing voltage (with external triggering). As in
the Burst Average mode, this command has no effect in conjunction with trigger sources BUS, HOLD
and IMMediate.
The query yields
1 for POSitive,
2 for NEGative.
Default setting: POSitive
TRIGger:SOURce[?] HOLD | IMMediate | INTernal | BUS | EXTernal |
TRIGger:SOURce sets the trigger source.
HOLD: Triggering only with command TRIGger:IMMediate.
IMMediate: Automatic triggering without explicit event.
INTernal: Triggering by the measurement signal. Relevant parameters:
TRIGger:LEVel, TRIGger:SLOPe, TRIGger:DELay, TRIGger:HYSTeresis,
TRIGger:HOLDoff.
BUS: Triggering with command *TRG or TRIGger:IMMediate.
EXTernal: Triggering via the hardware trigger bus, e.g. via the R&S NRP-Z3 USB adapter
or the R&S NRP base unit. Relevant parameters:
TRIGger:SLOPe, TRIGger:DELay, TRIGger:HOLDoff.