EasyManua.ls Logo

Rohde & Schwarz R&S NRP-Z91 - Sense:sgamma:magnitude[?] 0.0 to 1.0; Sense:sgamma:phase[?] -360.0 to 360.0

Rohde & Schwarz R&S NRP-Z91
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S NRP-Z91 Remote Control - Commands
1168.8579.12 6.13 E-3
Fig. 6-2 Correction of interactions between the power sensor, the signal source, and the s-
parameter device
The query yields
1 for OFF,
2 for ON.
Default setting: OFF
SENSe:SGAMma:MAGNitude[?] 0.0 to 1.0
SENSe:SGAMma:MAGNitude defines the magnitude of the complex reflection coefficient of the signal
source. A value of 0.0 corresponds to ideal match and a value of 1.0 to total reflection.
The query yields the set magnitude.
Default setting: 0.0
SENSe:SGAMma:PHASe[?] –360.0 to 360.0
SENSe:SGAMma:MAGNitude defines the phase angle (in degrees) of the complex reflection coefficient
of the signal source.
The query yields the set phase angle.
Default setting: 0.0 [°]

Table of Contents