Remote Control - Commands R&S NRP-Z91
1168.8579.12 6.14 E-3
SYSTem
With the aid of the SYSTem system, administrative device settings can be defined and queried. This
includes detailed information on the sensor and its initialization and the transfer of available commands
and their parameter limits.
Table 6-4 Commands of the SYSTem system
Command Parameter Unit Remarks
SYSTem
:INFO? [Item] Query only
:INITialize No query
MINPower? W Query only
RUTime[?] 0.0 to 10.0 s
SUTime[?] 0.0 to 10.0 s
:TRANsaction
:BEGIN: No query
:END No query
SYSTem:INFO? [Item]
SYSTem:INFO? yields a string containing information that is more detailed than the identification string
delivered by the sensor as a response to *IDN?. If no Item is specified, the response string is a
sequence of entries in the form Item:Information-String separated by CR and LF (in C notation: \r\n).
With the Item optionally appended to the command, the entry for the required Item can be queried. The
response string is zero-terminated, i.e. its end identification is a zero byte (in C notation: \0).
Table 6-5 Meaning of Item in the SYSTem:INFO? command
Item Information string Remarks
"MANUFACTURER" "Rohde & Schwarz GmbH & Co. KG" Manufacturer
"TYPE" "NRP-Z91" Type designation
"STOCK NUMBER" "1168.8004.02" Material number
"SERIAL" "<serial number>" 6-digit serial number
"HWVERSION" "000000000" Hardware version (standard)
"HWVARIANT" "000000000" Hardware model (standard)
"SW BUILD" "<build number>" Version number of sensor firmware

Table of Contents