R&S NRP-Z91  Remote Control - Commands 
1168.8579.12 6.5  E-3 
SENSe (Sensor Configuration) 
The sensor is configured by means of the commands of the groups SENSe and TRIGger.
Table 6-2  Commands of the SENSe system 
Command  Parameter  Unit  Remarks 
SENSe    
:AVERage    
:COUNt[?]  1 to 65536     
:AUTO[?]  OFF | ON | ONCE     
:TYPE[?]  RESolution | NSRatio     
:MTIMe[?] 0.01 to 999.99  s  
:NSRatio[?]  0.0 to 1.0  dB   
:RESolution[?] 1 to 4     
:RESet   No query 
:STATe[?] OFF | ON    
:TCONtrol[?] MOVing | REPeat    
:CORRection    
:DCYCle[?] 0.001 to 99.999  %  
:STATe[?] OFF | ON    
:OFFSet[?] –200.0 to 200.0  dB  
:STATe[?] OFF | ON    
:SPDevice:STATe[?] OFF | ON     
:FREQuency[?] 9.0e3 to 6.0e9  Hz  
:FUNCtion[?] “POWer:AVG“    
:POWer    
:AVG    
:APERture[?] 10.0e-6 to 0.3  s  
:BUFFer    
:SIZE[?] 1 to 1024    
:STATe[?] OFF | ON