R&S NRP-Z91  Remote Control - Commands 
1168.8579.12 6.3  E-3 
SCPI Commands 
The sensor R&S NRP-Z91 is controlled via the groups of commands 
• CALibration (zeroing) 
• SENSe (measurement configurations) 
• SYSTem 
• TRIGger 
• SERVice. 
 
CALibration 
Table 6-1  Commands of the CALibration system 
Command  Parameter  Unit  Remarks 
CALibration 
:DATA[?]  <calibration data set as definite length block> 
:LENGth?  Bytes  Query only 
:ZERO 
:AUTO[?]  OFF | ON | ONCE 
CALibration:DATA[?]   <calibration data set as definite length block>
CALibration:DATA is used for writing a calibration data set in the flash memory of the sensor. 
 
The query yields the calibration data set currently stored in the flash memory as a definite length block.
CALibration:DATA:LENGth? 
CALibration:DATA:LENGth? yields the length in bytes of the calibration data set currently stored in the 
flash memory. Programs that read out the calibration data set can use this information to determine the 
capacity of the buffer memory required.