R&S ZVH Power Meter (R&S ZVH-K9)
Performing Pulse Power Measurements (R&S ZVH-K29)
Operating Manual 1309.6946.12 - 09 169
5.4 Performing Pulse Power Measurements (R&S ZVH-
K29)
When you equip the R&S ZVH with firmware option R&S ZVH-K29, and connect one of
the wideband power sensors available from Rohde & Schwarz (R&S NRP-Z81, -Z85 or
-Z86), you can perform pulse power measurements with your R&S ZVH.
Support of the application for R&S ZVH models
For R&S ZVH with a serial number <115331 (model .04) or serial number <115239
(model .08), it’s mandatory to use Y adapter cable R&S FSH-Z129 (order no.
1304.5887.00).
Make sure to connect this cable to the USB port and the AUX port on the left side of
the R&S ZVH.
Like the normal power meter application, the pulse power application measures the
power of the whole signal in the frequency range of the (wideband) power sensor.
Press the MODE key.
Press the "Power Meter" softkey.
Connecting the power sensor
You can connect the wideband power sensors to the USB port of the R&S ZVH. For
more information see "Connecting a Power Sensor" on page 159.
The measurement starts as soon as the power sensor is connected.
Numerical result display
Press the MEAS key.
Press the "Meas Mode" softkey.
Select the "Average" menu item.
The layout and contents of the numerical result display are the same as those
described in chapter "Using a Power Sensor" on page 157.
Graphical result display (Power vs Time)
The graphical representation of the results is a special feature only available with the
firmware option R&S ZVH-K29.
Press the MEAS key.
Press the "Meas mode" softkey.
Select the "Power vs Time" menu item.