56 Configuration, Startup and Operation
Section 3: Configuration, Startup and Operation Reference Manual
January 2019 00809-0100-4892
5. From the CALIBRATION menu, select TOLERANCE CHECK.
6
. Press the RIGHT arrow key to change the value from NO to YES or YES to NO.
No – Tolerance Check active
Yes – Tolerance Check inactive
Use the stylus on the screen to press ENTER when finished.
7. Use the stylus on the screen to select SEND to update the 6888Xi.
3.7.5 Alarm Relay Output Configuration
The 6888Xi has two dry contact Form-C alarm relay output signals that can be configured in
multiple modes through the 6888Xi keypad display or the 375/475 Field Communicator. A list
of modes is shown in Table 3-3. Each alarm relay output can be configured separately and one
or all modes can be turned on in any combination.
If the 6888Xi is configured with the optional Flame Safety Interlock or for Traditional
Architecture, Alarm 2 is preconfigured with "Heater Relay" and prewired to either the AC Relay
Board or DR Board. In this condition the relay configuration cannot be changed to any other
setting. If changes are attempted to be made, the 6888Xi will respond with an “Access
Restricted” warning. If the 6888Xi is not configured with the optional Flame Safety Interlock,
"Heater Relay" is not valid and cannot be chosen for Alarm 2.
Configuring the Alarm Relays Using the 6888Xi Keypad/Display
1. Press the MENU button repeatedly until the IOB1 Probe 1 menu displays.
2. From the main menu, select DETAILED SETUP then press ENTER.
3. From the DETAILED SETUP menu, select RELAY then press ENTER.
4. From the RELAY menu, select RELAY 1 then press ENTER.
5. From the RELAY 1 menu, use the UP and DOWN arrow keys to select the mode and the
ENTER key to toggle the value from ON to OFF or OFF to ON. Relay modes are as follows:
Unit Alarm
Low O2
Cal Recommended
In Calibration
6. Press ENTER key to accept settings. When questioned, use the UP and DOWN arrow keys to
select YES or NO to save configuration. Press ENTER to continue.
7. From the RELAY menu, select RELAY 2 then press ENTER.
8. From the RELAY 2 menu, use the UP and DOWN arrow keys to select the mode and the
ENTER key to toggle the value from ON to OFF or OFF to ON. Relay modes are as follows:
Unit Alarm
Low O2
Cal Recommended
In Calibration
Heater Relay
9. Press LEFT arrow key. When questioned, use the UP and DOWN arrow keys to select YES or
NO to save configuration. Press ENTER to continue.
10. Press the EXIT key to return to the main menu.
11. Repeat steps 1 through 10 for PROBE 2 if configured for dual channel.
Configuring the Alarm Relays Using Field Communicator in Manual Configure Mode