Configuration, Startup and Operation 67
Reference Manual Section 3: Configuration, Startup and Operation
00809-0100-4892 January 2019
Check feature is active.
A
utoCal No Resp – No communication or loss of communication to the SPS4001B or IMPS.
AutoCal OutofSync – Loss of communication synchronization to the SPS4001B or IMPS.
AutoCal Abort – The SPS4001B or IMPS aborted the calibration cycle due to a low gas bottle
indication.
WarmUp Abort – Attempted to perform a calibration during warmup.
Alarm Abort – Another alarm occurred and caused the calibration cycle to abort.In the
event of a calibration failure, perform the following steps. Otherwise, proceed to the next
step.
a. From the RESULT menu, press HOME to return to the main menu.
b. From the main menu, select SERVICE TOOLS.
c. From the OVERVIEW menu, select ALERTS.
d. From the ALERTS menu, select ACTIVE ALERTS.
e. The ACTIVE ALERTS menu should contain an A: CALIBRATION FAILED alarm. Press the
left arrow key to return to the ALERTS menu.
f. From the ALERTS menu, select DEVICE STATUS.
g. From the DEVICE STATUS menu, select ACKNOWLEDGE.
h. From the ACKNOWLEDGE menu, select ACK CALIBRATION FAILED. When the process is
complete, the screen will return to the ACKNOWLEDGE menu.
17. If the Calibration Acknowledge parameter is ON, perform the following steps. Otherwise,
proceed to the next step.
a. Press HOME to return to the main menu.
b. From the main menu, select SERVICE TOOLS.
c. From the OVERVIEW menu, select ALERTS.
d. From the ALERTS menu, select ACTIVE ALERTS.
e. The ACTIVE ALERTS menu should contain an M: CALIBRATION CHANGED alarm. Press
the left arrow key to return to the ALERTS menu.
f. From the ALERTS menu, select DEVICE STATUS.
g. From the DEVICE STATUS menu, select ACKNOWLEDGE.
h. From the ACKNOWLEDGE menu, select ACK CALIBRATION.
i. The next screen shows the CURRENT CAL and NEW CAL slope and constant values.
Press OK to continue.
j. Use the UP and DOWN arrow keys to change the value from NO to YES or YES to NO.
Press ENTER to continue. If YES is selected, the Field Communicator will respond CALI-
BRATION CONSTANTS ACCEPTED. ALERT CLEARED. Press OK to return to the
ACKNOWLEDGE menu.
18. If the Diffuser Warning Function is active, perform the following steps.
a. Press HOME to return to the main menu.
b. From the main menu, select SERVICE TOOLS.
c. From the OVERVIEW menu, select ALERTS.
d. From the ALERTS menu, select ACTIVE ALERTS.
e. The ACTIVE ALERTS menu should contain an A: DIFFUSER WARNING alarm. Press the
left arrow key to return to the ALERTS menu.
f. From the ALERTS menu, select DEVICE STATUS.
g. From the DEVICE STATUS menu, select ACKNOWLEDGE.
h. From the ACKNOWLEDGE menu, select ACK DIFFUSER WARNING. When the process is
complete, the screen will return to the ACKNOWLEDGE menu.
Aborting O
2
Calibration with 6888Xi and Field Communicator
1. From the calibration prompt screen, press the OK button to return to the SELECT ACTION