EasyManua.ls Logo

RST VW2106 - 3 CONFIGURATION; 3.1 SINGLE FIELD INSTRUMENT CONNECTION

RST VW2106
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VW2106 Vibrating Wire Readout Manual
ELM0042M
RST Instruments Ltd.
Page 4
NOTE: THE AUTO POWER OFF FEATURE IS ALWAYS ACTIVE TO CONSERVE BATTERY
LIFE AND CANNOT BE DISABLED.
4. The next screen prompts the user to select the Auto Backlight Off (Figure 2-4).
Use the up/down arrows to adjust the desired time and press Enter.
FIGURE 2-4 AUTO BACKLIGHT OFF
5. The next screen prompts the user to select the Speaker On time. Use the
up/down arrows to select the desired time and press Enter. The unit will return to
the previous menu.
FIGURE 2-5 SPEAKER ON TIME
NOTE: KEEP THE UNIT OFF WHEN NOT IN USE TO PRESERVE BATTERY LIFE.
3 CONFIGURATION
The VW2106 may be connected to either single field instruments or multi-channel
instruments.
3.1 SINGLE FIELD INSTRUMENT CONNECTION
The following instructions outlined below to connect single field instruments:
1. Match each wire to its respective terminal post. The posts have colour coded
labels for quick reference (Figure 3-1).

Table of Contents