DESCRIPTION OF THE INSTRUMENT
Carl Zeiss Technical data Axio Lab.A1
22 430037-7144-001 04/2013
Light sources
LED transmitted light
Power consumption ............................................................................................................. max. 3 W
Adjustment of light source ................................................................... continuous approx. 0.5 to 12 V
Halogen lighting transmitted light
Power consumption ........................................................................................................... max. 35 W
Halogen lighting reflected light
Power consumption ................................................................................................................... 50 W
Adjustment of light source ................................................................... continuous approx. 0.5 to 12 V
LED lighting reflected light fluorescence with replaceable LED modules
Wavelengths optional .............................................. 365, 380, 455, 470, 505, 530, 590, 615, 625 nm
.......................................................................................................... or neutral white (540 - 580 nm)
LED classification ...................................................................................... LED Risk Group 2 to IEC 62471
Axio Lab.A1:
Stand with manual stage focusing
Coarse focusing ............................................................................................ approx. 4 mm/revolution
Fine focusing ................................................................approx. 0.4 mm/revolution; 4 μm scale interval
Lifting range ............................................................................................................................ 30 mm
Height stop ............................................................................................................... pre-set at works
Optional condenser 0.9/1.25 H with or without modulator disk for
................................................................................... Brightfield, darkfield and phase contrast 1, 2, 3
Manual objective change ............................................................... by nosepiece, 4x H Pol or 5x H D, M27
Manual reflector module change .............................................................................. by reflector turret 4x