Section 3
Revised 3/30/2010 3.9
Setting the Slowdown Position
To change the pressure applied to the cover glass
during coverslipping, use this function (Figure 3-U).
NOTE: This menu item is available only when the system
is on standby or paused or when the system check screen
is displayed.
Quality coverslipping can be achieved by pressing on
the cover glass with appropriate force. If air bubbles are
present or mounting medium seeps out from the sides,
take the following actions:
If air bubbles are present, raise the slowdown
position (increase the pressure).
If mounting medium seeps out from the sides,
lower the slowdown position (decrease the
pressure).
Figure 3-U
Figure 3-V: Slowdown Position Setting Screen
1. Press the MENU key twice.
2. Press the < and > keys to highlight 1, “Slowdown
position, and then press the ENTER key. The
slowdown position setting screen appears (Figure 3-V).
3. Pressing the < key decreases the value, while
pressing the > key increases the value.
4. Press the ENTER key to save the selected setting. To
return to the menu screen without saving the change,
press the EXIT key.
<Slowdown pos> 60mm
005
[<][>] to select
[ENTER] to save

Other manuals for Sakura Tissue-Tek Glas g2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Glas g2 and is the answer not in the manual?

Sakura Tissue-Tek Glas g2 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Glas g2
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals