Section 4
Revised 3/30/2010 4.19
3. Press the ENTER key on the numeric keypad to
confirm the coverslipping program, and confirm that
the program number has been changed (Figure 4-
EE). Set (save) a coverslipping program for each
staining program. When setting a coverslipping
program, make sure the applicable program is already
programmed on the coverslipper side.
Figure 4-EE
4. Open the Prisma loading station access door, place
baskets in the specified loading station, and close the
door (Figure 4-FF).
Figure 4-FF
5. After confirming the staining program and the loading
station where the baskets have been placed, press
the START key to start the process (Figures 4-GG &
4-HH).
Figure 4-GG
Figure 4-HH
WARNING: MAKE SURE THE SLIDES ARE EVENLY
STORED IN THE BASKET. IF THE SLIDES ARE
STORED UNEVENLY, THE BASKET MOVEMENT IS
AFFECTED DURING THE STAINING OPERATION.
TAKE SPECIAL CARE WHEN THERE ARE ONLY A
SMALL NUMBER OF SLIDES. STORE THEM
CAREFULLY TO PREVENT THE BASKET FROM
TILTING.

Other manuals for Sakura Tissue-Tek Glas g2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Glas g2 and is the answer not in the manual?

Sakura Tissue-Tek Glas g2 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Glas g2
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals