5-18
PCU-110 Operation Manual
RQS (Request Service Bit):
RQS is set when a Service Request is issued.
ESB (Event Status Bit):
ESB is set when any of the bits of the Standard Event Status
Register (SESR) is set.
MAV (Message Available Bit):
MAV is set when output data is prepared.
3) Service Request Enable Register (SRER)
An 8-bit register to set Enable/Disable of each bit of the Status Byte
Register. If a bit of the SRER is set to 1, the corresponding bit of the
STBR is set to Enable.
4) Standard Event Status Register (SESR)
An 8-bit register to indicate the standard status of the device.
PON (Power ON):
Bit is set when power is turned on.
CME (Command Error):
CME is set when a command has a syntax error or is
misspelled.
EXE (Execution Error):
EXE is set when a command is not available for the device, or a
command cannot be executed in the current device condition.
DDE (Device Denition Error):
Not used
QYE (Query Error):
QYE is set when an attempt is made to read data from the
output queue while the output queue is empty.
OPC (Operation Completion):
OPC is set when all pending operations are completed.
5) Standard Event Enable Register (SEER)
SEER is an 8-bit register to set Enable/Disable of each bit of the
SESR. If a bit of the SEER is set to 1, the corresponding bit of the
SESR is set to Enable.
PON
Not used
CME EXE DDE QYE
Not used
OPC