Polarization Control Unit PCU-110
9-1
9
Troubleshooting
Fault condition Cause Action
No power Cord is unconnected. Connect cord properly.
No fuse or fuse is blown out. Open the fuse box lid and
replace fuse.
No light output End face of the optical connector
is dirty.
Clean the end face of the optical
connector.
Optical connector is not connected
properly
Check the connecting state of the
optical connector.
GPIB communication
is not possible
GPIB address is not set correctly. Set GPIB address correctly.
(see section 5-1)
The delimiter of this device does
not match with the delimiter of the
control software.
Match the delimiters.
(see section 5-1)
GPIB cable is not connected
properly.
Check the connection of the
cable. (see section 5-1)
LAN communication is
not possible
IP address is not set correctly Set IP address correctly. (see
section 5-2)
LAN cable is not connected
correctly.
Check the connection of the cable.
(see section 5-2)
USB communication is
not possible
USB driver is not installed on the
computer properly.
Install the USB driver using the
bundled removable disk.
(see section 5-3)
USB cable is not connected
properly.
Check the connection of the cable.
(see section 5-3)
If after checking the above list, and the device is still not operating correctly, please