26
Model AZ212 AZ6101 AZ4101 AZ2101 AZ601
Weighing capacity g 210 6,100 4,100 2,100 610
Readability g 0.01 0.1 0.1 0.1 0.1
Tare range (subtractive) g 210 6,100 4,100 2,100 610
Repeatability <±g 0.015 0.1 0.1 0.1 0.1
Linearity <±g 0.03 0.3 0.3 0.3 0.3
Operating temperature range +10… +30°C (50° to 86°F)
Allowable ambient 
operating temperature +5… +40°C (41°F to 104°F)
Sensitivity drift
within +10…+30 °C <±/K 1·10
–5
0,5 ·10
–5
1·10
–5
2·10
–5
5·10
–5
Response time (average) s 2.5 2221.5
Adaptation to 
ambient conditions By selection of 1 of 4 optimized filter levels
Display update (depends 
on the filter level selected) s 0.2–0.8
External calibration weight g 100 (M1) 5000 (F2) 2 (F2) 1 (M1) 0.5 (M1)
(of at least accuracy class) lb 0.2  10 4 2 1
Net weight, approx.  kg 1.4 1.7 1.7 1.7 1.7
Pan size mm 116 d 174+143 174+143 174+143 174+143
Dimensions (B+T+H) mm 200+270+70
AC power source/ 
power requirements  AC adapter 230 V or 115 V, +15% to –20%
Frequency 48–60 Hz
AC power source/
direct current V 10–20
Power consumption 
(average) W 0.75
Hours of operation with 
the YRB08Z rechargeable 
battery pack h 25