Changing the Reference Percentage
You can change the reference percentage in cycles. Choose from the following
settings: 5, 10, 20, 50 and 100.
To activate the Hold down v (8) for more than 2 seconds until
change function: “rEF 100 %” appears on the display
To change the setting: Briefly press v
To store the percentage: Hold down v for more than 2 seconds to store
this value permanently (in the non-volatile memory)
after you turn off the power
This setting is not canceled by the reset code 9 – – 1°!
Storage Parameter for the Reference Weight/Value
The reference weight/value is stored… Code
with full accuracy according to the internal resolution 3 5 1*
according to the display accuracy 3 5 2
When you exit the reference storage mode, “Err 22” may briefly be displayed.
This means that the new reference value has been stored.
Toggling between the Readout in Percent (%) and Weight (g)
After placing the sample on the weighing pan, you can toggle between the
readout in percent and the respective weight readout in grams by pressing
the v key after you have stored the reference value. To clear the readout in
percent, use the c key.
Display Parameter for Readouts in Percent
The following display parameters can be set for readouts in percent:
The readout in percent is displayed… Code
without a decimal place 3 6 1
with one decimal place 3 6 2*
with two decimal places 3 6 3*
with three decimal places 3 6 4
If the weight stored is too light to be displayed, the number of decimal places
is automatically decreased.
* = factory setting; depends on the balance model in some cases
43