Data Printout/Record for Application Programs
For application programs, reference data (parameters) can be included
in the printout/record.
To have a data record printed out, perform the following:
– Press v to output the printout heading and reference data (the reference data
will be stored at the same time)
or
– Press p to output the printout heading and the first value
If you input and store new reference values while an ISO/GLP-compliant record is
being printed out, the new reference data will be output. If you enter different
data before generation of an ISO/GLP record has started, the printout heading
and the reference data will automatically be printed once you press p. Then the
measured value will be output.
– Press p to output weighing data
– Press c to end printout generation (generation of an ISO/GLP printout also
ends once an adjustment or calibration operation has been started)
– Then press c to clear the reference data for the application programs
The printout can have the following lines:
--------------------
MC1 - Sartorius : Balance family and manufacturer
Model BP211D : Balance model
S/N 040500048 : Serial number of the balance
ID : Space for entering the workstation or operator ID
--------------------
Date : 21-Mar-97 : Current date
Start: 14:04:12 : Time at which the application started
Ser. : : Space for entering a project number
nRef + 10 pcs : Reference data (in this case,“counting” –
see also page 45)
wRef + 0.13400 g
Qnt + 500 pcs : Measured values (in this case
“calculated piece count”)
:
:
End : 14:13:07 : End of application
Name : : Space for signature of the operator responsible
--------------------
53