ISO/GLP-compliant Printout or Record
Application:
Use of the balance as a test and measuring instrument in quality assurance
systems in compliance with the requirements of ISO, GLP, GMP and EN (European
Standards) in which proof of the balance’s accurate performance is required.
The balance can record all completed calibration or adjustment operations and
print out data in compliance with the requirements of Good Laboratory
Practice (GLP). The balance, interfaced with a data printer or a computer, creates
a document that records the date, time, serial number and model number,
making it possible to clearly trace data to the balance that generated it and the
time at which it was generated.
Select the ISO/GLP-compliant printout or record mode by setting the respective
code in the balance operating menu:
ISO/GLP-compliant printout/record mode Code
Off 8 10 1*
Only for adjustment/calibration and linearization functions 8 10 2
Always on (e.g., for adjustment/calibration and linearization
functions, application programs, weight readouts) 8 10 3
The following menu code setting must be selected in order to obtain an ISO/GLP-
compliant printout/record:
With data ID code 7 2 2
!
Important Note
ISO/GLP-compliant printouts/records will not be generated if the
factory setting, code 7 2 1, is selected. In addition, do not select the
“Auto print” data output parameter (code 6 1 4 or 6 1 5).
Operating the Balance with a Device That Has an ISO/GLP Printing
or Recording Capability
You can connect a special Sartorius Data Printer to Basic
plus
balances for generating
ISO/GLP-compliant printouts
(order no. YDP03-0CE).
To generate ISO/GLP-compliant records with a computer, you will need
special software. For information on writing this software, please ask Sartorius for
a detailed description.
* = factory setting