Storage Threshold for the Automatic Start Mode
To obtain an added measure of reliability in the automatic mode, in order to
avoid an “incorrect start,” a weight value must correspond to a minimum load
of 100 display increments.
Practical Example: Animal Weighing in the Automatic Start Mode
Menu code settings used in the example:
Function Code
Animal weighing 2 1 7
Delay start until dif. is average 3 7 2*
Automatic start mode 3 8 2*
Automatic output
of all parameters 7 1 2
With data ID code 7 2 2
Application: Automatic weighing of animals based on10 subweighing operations
Step/Key Readout Data Output/Printout
Place animal weighing bowl
on balance
c, t
0.00 g
Place 1st animal in bowl
v
10 g
mdef + 10
465.20 g
x-Net + 465.20 g
Remove 1st animal
Place 2nd animal in bowl
10 g
mdef + 10
388.53 g
x-Net + 388.53 g
Repeat above steps
for all animals
The data ID codes mean:
mdef + 10 defined number of subweighing operations
x-Net + 401.18 g calculated average
* = factory setting
As an alternative, you can select
this output parameter:
Other delay start 3 7 x