For data communication with a software handshake, “XON” must be sent
by a device when it is turned on in order to enable another on-line device to
exchange data.
Sequence:
Balance Receiving device
byte
byte
byte
<XOFF>
byte
(Once <XOFF> has been transmitted, a maximum
byte
of 14 bytes can still be received.)
::
:Pause:
::
::
<XON>
byte
byte
Transmitting device:
The importance of handshake control for data transmission becomes
especially apparent
when the continuous automatic data output parameter is defined
when data output is controlled by application programs
Once <XOFF> has been received, it prevents further transmission of characters.
When <XON> is received, it re-enables the transmitting device to send data.
The transmitting device is always enabled for sending data after it has been
switched on.
If data transmission is interrupted by the control line (CTS) or the command
<XOFF> while a data block is being output from an application program (only for
printing a section of text with several lines of data), the readout will be locked
into the display at the same time.
Data output will be blocked until the interface receives an enabling signal.
64

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius BP 3100 P-0CE and is the answer not in the manual?

Sartorius BP 3100 P-0CE Specifications

General IconGeneral
BrandSartorius
ModelBP 3100 P-0CE
CategoryScale
LanguageEnglish

Related product manuals