Installation Instructions
Ambient Conditions
When choosing a location to set up your balance,
observe the following:
– Set up the balance on a stable, even surface
(benchtop or floor), or place it on a wall console
(see “Accessories”)
– Avoid placing the balance in close proximity
to a heater or otherwise exposing the balance
to extreme heat or to direct sunlight
– Protect the balance from drafts that come
from open windows and doors
– Avoid exposing the balance to extreme vibrations
during weighing
– Protect the balance from aggressive
chemical vapors
– Do not operate the balance in a hazardous
area/location
Do not expose the balance to extreme moisture over
long periods. Moisture in the air can condense on
the surfaces of a cold balance whenever it is brought
to a substantially warmer place. If you transfer the
balance to a warmer area, make sure to condition it
for about 2 hours at room temperature, leaving it
unplugged from AC power. Afterwards, keep the
balance connected continuously to AC power.
13