Data Printout/Record (ISO/GLP-compliant)
To have a data record printed out, perform the following:
Press p to output the printout heading and the first value (after you have turned
on the balance or cleared a function by pressing c)
Press p to output additional data
To end printout generation and recording of data, press c
Generation of an ISO/GLP-compliant printout or record is also ended when
an adjustment or calibration operation is started.
The printout can have the following lines:
--------------------
MC1 - Sartorius : Balance family and manufacturer
Model BP211D : Balance model
S/N 040500048 : Serial number of the balance
ID : Space for entering the workstation or operator ID
--------------------
Date : 21-Mar-97 : Current date
Start: 11:00:30 : Time at which the application started
Ser. : : Space for entering a project number
N+5.45390g: Measured weights
N+24.34586g
N+63.23450g
End : 11:15:55 : End of application
Name : : Space for signature of the operator responsible
--------------------